Description
Testo 330-2G LX Commercial / Industrial Analyzer with Bluetooth and NOx - Kit 2
The Testo 330-2G with Long-Life sensor technology marks a milestone in combustion analysis. In combustion analysis, proper set up and maintenance are critical to safe and efficient operation. The Testo 330-2G makes it simple. No more guesswork! Graphic boxes and trending curves display when your measurement is on target or off. The graphic display provides easy analysis of measurement data, providing more functions than ever before including testing for pipe commissioning and differential pressure. The interface for burner control modules makes this the most versatile tool in combustion tuning and testing.
The Testo 330-2G has the longest service life available in a combustion analyzer. Improved materials and construction details provide you with reliability and the lowest cost-of-ownership. It comes with the longest warrantee in the market with an expected sensor life span of up to 6 years (O2 and CO sensors). Rest assured the Testo 330-2G LX is an essential tool for combustion analysis.
Features:
- Intuitive graphic display
- Simple graphics alert you when combustion parameters are out of range
- Long Life Sensors with over-range protection
- Measure CO up to 30,000 ppm(with standard dilution system)
- Instrument diagnostics and quick leak check at the touch of a button
- On-board memory to ½ million readings
- Optional EasyHeat Software for data management
- 20 selectable fuels with 10 user defined fuels
- Calculations: CO2, efficiency, excess air, CO air free
- Differential pressure
- Differential temperature(with optional probes)
- Gas pipe testing: pressure drop, pipe commissioning, and more
- Gas leak with optional probe
- Temperature range: -40 to 2,192 °F
- USB connectivity and more
Specs
| Differential Pressure | |
| Measuring range | 0 to +120 InH2O / 0 to +300 hPa |
| Accuracy | ±0.2 InH2O (0.0 to 20.0 InH2O) ±0.5 hPa (0.0 to 50.0 hPa) ±1 % of mv (20.0 to 40.0 InH2O) ±1 % of mv (50.1 to 100.0 hPa) ±1.5 % of mv (Remaining Range) |
| Resolution | 0.1 InH2O / 0.1 hPa |
| Flue gas O2 | |
| Measuring range | 0 to 21 Vol.% |
| Accuracy | ±0.2 Vol.% |
| Resolution | 0.1 Vol.% |
| Reaction time t90 | < 20 sec |
| Flue gas CO | |
| Measuring range | 0 to 8000 ppm |
| Accuracy | ±10 ppm or ±10 ppm (0 to 200 ppm) ±20 ppm or ±5 ppm (201 to 2000 ppm) ±10 % of mv (2001 to 8000 ppm) |
| Resolution | 1 ppm |
| Reaction time t90 | < 60 sec |
| Flue gas COlow | |
| Measuring range | 0 to 500 ppm |
| Accuracy | ±2 ppm (0 to 39.9 ppm) ±5 % of mv (40 to 500 ppm) |
| Resolution | 0.1 ppm |
| Reaction time t90 | < 40 sec |
| CO determination (with H2-compensation), automatic dilution | |
| Measuring range | 0 to 30000 ppm |
| Accuracy | ±100 ppm (0 to 1000 ppm) ±10 % of mv (1001 to 30000 ppm) |
| Resolution | 1 ppm |
| Flue Gas NO | |
| Measuring range | 0 to 3000 ppm |
| Accuracy | ±5 ppm (0 to 100 ppm) ±5 % of mv (101 to 2000 ppm) ±10 % of mv (2001 to 3000 ppm) |
| Resolution | 1 ppm |
| Reaction time t90 | < 30 sec |
| Flue gas NOlow | |
| Measuring range | 0 to 300 ppm |
| Accuracy | ±2 ppm (0 to 39.9 ppm) ±5 % of mv (40 to 300 ppm) |
| Resolution | 0.1 ppm |
| Reaction time t90 | < 30 sec |
| Flue Gas Draft | |
| Measuring range | -4.00 to +16 InH2O / -9.99 to +40 hPa |
| Accuracy | ±0.01 InH2O or ±2 InH2O (-0.50 to +0.24 InH2O) ±0.02 hPa or ±5 hPa (-0.50 to +0.60 hPa) ±0.01 InH2O (+0.24 to +1.20 InH2O) ±0.03 hPa (+0.61 to +3.00 hPa) ±1.5 % of mv (+3.01 to +16.00 InH2O) ±1.5 % of mv (+3.01 to +40.00 hPa) |
| Resolution | 0.01 InH2O / 0.01 hPa |
| Temperature | |
| Measuring range | -40° to 2192 °F / -40 to +1200 °C |
| Accuracy | ±0.9 °F (32° to 212.0 °F) / ±0.5 °C (0 to +100.0 °C) ±0.5 % of mv (Remaining Range) |
| Resolution | 0.1 °F (-40° to 1831.8 °F) / 0.1 °C (-40 to +999.9 °C) 1.0 °F (> +1832 °F) / 1 °C (> +1000 °C) |
| Efficiency (calculated) | |
| Measuring range | 0 to 120 % |
| Resolution | 0.1 % |
| Flue Gas loss (calculated) | |
| Measuring range | 0 to 99.9 % |
| Resolution | 0.1 % |
| Flue Gas CO2 calculation (calculated from O2) | |
| Measuring range | 0 to CO2 max (Display range) |
| Accuracy | ±0.2 Vol.% |
| Resolution | 0.1 Vol.% |
| Reaction time t90 | < 40 sec |
| Pressure measurement | |
| Measuring range | 0 to +120 InH2O / 0 to +300 hPa |
| Accuracy | ±0.2 InH2O (0.0 to 20.0 InH2O) / ±0.5 hPa (0.0 to 50.0 hPa) ±1 % of mv (50.1 to 40.0 InH2O) / ±1 % of mv (50.1 to 100.0 hPa) ±1.5 % of mv (Remaining Range) |
| Resolution | 0.1 InH2O / 0.1 hPa |
| Flue gas CO (without H2-compensation) | |
| Measuring range | 0 to 4000 ppm |
| Accuracy | ±20 ppm (0 to 400 ppm) ±5 % of mv (401 to 2000 ppm) ±10 % of mv (2001 to 4000 ppm) |
| Resolution | 1 ppm |
| Reaction time t90 | < 60 sec |
| Ambient CO | |
| Measuring range | 0 to 500 ppm |
| Accuracy | ±5 ppm (0 to 100 ppm) ±5 % of mv (> 100 % of mv) |
| Resolution | 1 ppm |
| Reaction time | Approx. 35 sec |
| with CO probe 0632 3331 | |
| Ambient CO2 | |
| Measuring range | 0 to 1 Vol.% 0 to 10000 ppm |
| Accuracy | ±75 ppm or ±3 % of mv (0 to 5000 ppm) ±150 ppm or ±5 % of mv (5001 to 10000 ppm) |
| Reaction time | Approx. 35 sec |
| with ambient CO2 probe 0632 1240 | |
| Gas leak measurement for combustible gases (via gas leak detection probe) | |
| Measuring range | 0 to 10000 ppm CH4 / C3H8; Display range |
| Accuracy | Signal optical display (LED) audible signal via buzzer |
| Reaction time t90 | < 2 sec |
| with gas leak detection probe 0632 3330 | |
| General technical data | |
| Protection class | IP40 |
| Warranty | Probe: 48 months Gas sensors (O2, CO) 48 months NO sensor 24 months COlow sensor 24 months Thermocouple 12 months rechargeable battery: 12 months |
| Display size | 240 x 320 pixels |
| Display function | Colour graphic display |
| Power supply | Rechargeable battery pack 3.7 V / 2.6 Ah; Mains unit 6 V / 1.2 A |
| Maximum | 500,000 readings |
| Storage temperature | -4° to 122 °F / -20 to +50 °C |
| Weight | 21 oz. (without rechargeable battery) / 600 g (without rechargeable battery) |
| Dimensions | 11 x 4 x 3 in. / 270 x 90 x 65 mm |
| Operating temperature | 23° to 113 °F / -5 to +45 °C |
Includes
- (1) Testo 330-2G LX Commercial / Industrial Analyzer with Bluetooth
- (1) 12" Flue Gas Probe with 7' Hose
- (10) Particle filters
- (1) O2 Sensor
- (1) H2 compensated CO Sensor
- (1) AC power adapter
- (1) Certificate of Conformity
- (1) Li-Ion Battery
- (1) Hard Case
Videos