U.E.i. is committed to providing a superior quality temperature tester and placed a strong emphasis on quality design, forward thinking innovation, and a systematic approach to product development, to ensure that we not only build a competitive thermometer line, but a superior one that creates real world solutions.

Features:

  • Dual Thermocouple inputs
  • Temp range -346 to +2192F (J)
  • Temp range -328 to +2498F (K)
  • Data logging
  • 9,999 Memory positions
  • USB interface (software included)
  • Thermocouple temperature offset
  • MIN, MAX, AVG and Hold
  • Dual Differential T1-T2
  • Probe Storage
  • IP67

Specs

Specification DT302
Measurement RangeJ-Type: -346° to +2192°F (-210° to +1200°C)
K-Type: -328° to +2498°F (-200° to +1370°C)
T-Type: -418° to +752°F (-250° to +400°C)
E-Type: -238° to +1832°F (-150° to +1000°C)
Display Resolution0.1°F/°C < 1000°
1.0°F/°C = 1000°
AccuracyJ, K, T, and E-Type; ±[0.1% +1.0°F (0.5°C)]
[Below -148°F (-100°C): add 0.2% of reading
for J, K, and E-Type; and 0.1% of reading for T-Type]
Temperature Coefficient0.01% of reading 0.1°F per °F(+0.05°C per °C)
for <+64°F (+18°C) or +82°F (+28°C)
[Below -148°F (-100°C): add 0.05% of reading
for J, K, and E-type; and 0.1% of reading for T-type]
Real Time Clock ToleranceAbout 1 second per day
Maximum Differential
Common Mode Voltage
1V (maximum voltage difference between any pair of inputs)
Temperature ScaleITS-90 (International Temperature Scale of 1990)
Applicable StandardsN.I.S.T. Monograph 175 revised to ITS-90
Dimension7.20” (H) x 3.70” (W) x 1.70” (D)
WeightApprox. 16.2 oz
Battery2 x LR03 (AAA) type 1.5V
CertificationCE
SafetyIEC 1010-1 (2001), UL 3111-1 (6, 1994)
EN 61010-1 (2001), CSA C22.2
No. 1010.1 (1992)
CAT IOver Voltage (Installation) CATEGORY I 
Pollution Degree 2 per IEC 1010-1
ComplianceIEC 60529: IP67 (Ingress Protection rating)
Water Immersion (3’ up to 30 min)
No ingress of dust (20mBar up to 8 hrs.)